+1 215-283-0860   |    This email address is being protected from spambots. You need JavaScript enabled to view it.

CNS 16 Input EEG Amplifier

CNS 16 Input EEG Amplifier

Continuous EEG Monitoring for your ICU

Collect continuous EEG along with all the other multimodal waveform and trends supported by the CNS Monitor.

This email address is being protected from spambots. You need JavaScript enabled to view it.

Features

  • Record up to 16 EEG electrode locations (including reference)
    Up to 6 of the inputs can be used as differential pairs to monitor EEG or other physiological signals (e.g. ECG, EOG, respiration, etc.)
  • Continuous independent impedance checking to verify electrode connection integrity (can be disabled by user)
  • Electrode indicator LEDs assist with setup and troubleshooting

Physical Specifications

  • 10.4 cm x 12.1 cm x 4.4 cm (4.08" x 4.75" x 1.75") W x L x H
  • ~ 0.34 kg (0.75 lbs.)

Interface Cable

  • MDportTM - USB electrical interfacing using a medical grade keyed latching connector
  • Cable Length: 5 meters (16.4 feet)

Electrode Connections

  • 19 Touchproof (DIN 42-802) safety connections

Amplifier Inputs

  • Reference Input
  • Ground Input
  • 11 dedicated referential electrode inputs
  • 2 inputs configurable as separate referential inputs or 1 differential pair
  • 2 inputs configurable as referential inputs or 2 differential pairs (when used with the dedicated referential jack associated with each input)
  • Selectable continuous impedance testing when the EEG Amplifier is connected to the Monitoring System’s computer. Measurement is accurate to 1 kΩ or 20%, whichever is greater. (Note: The impedance test signal can be turned off by the user.)

Electrical Specifications

  • Input Impedance: ~10 MΩ with impedance test circuit enabled (>20MΩ with the impedance test circuit disabled)
  • CMRR: ~100 dB at 50/60 Hz with no input imbalance with a 5kΩ impedance imbalance
  • Noise: Less than 2 μV p-p
  • Inter-channel Cross Talk: Less than 0.10% (-60db)
  • Gain: 33.5 ± 2.2 % before A/D converter
  • Bandwidth: 0.1 to 70 Hz to 70% response points (-3db)
  • Input Range: Up to 78 mV p-p full scale
  • A/D Conversion: 24 bit
  • Sampling Rate: 512 Hz